• evo视讯(中国)

    Opentrons Pipeline: DNA Extraction with the Mag-Bind Blood & Tissue DNA HDQ 96 Kit Tissue Protocol

    Thalía Silvestre

    Published: 2024-07-30 DOI: 10.17504/protocols.io.bp2l62rqrgqe/v1
    AI 解读
    浏览 1537

    Disclaimer

    DISCLAIMER – FOR INFORMATIONAL PURPOSES ONLY; USE AT YOUR OWN RISK

    The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment.

    Abstract

    This protocol is an automated pipeline to extract a full 96-well plate of DNA from tissue lysates.

    This protocol was developed and optimized for the following:

    Platform: Opentrons OT-2 automated pipetting robot

    Kit: Mag-B®nd‱ Blood & Tissue DNA HDQ Prefilled 96 Kit

    Recommended number of samples: 96

    Two ellutions: Firts ellution( 60μL ) , Second ellution ( 30μL)

    Steps

    Samples Lysis

    1.

    Preparate 96 samples lysis following the Mag-Bind® Blood & Tissue DNA HDQ 96 Kit Tissue Protocol:

    1.1.

    Mince up to 10 mg tissue and transfer to a 1.5 mL tubes (one for each sample).

    (Cut each 10 mg sample with sterilized cutters as small as possible)

    1.2.

    Add 250 μL TL Buffer to each sample.

    1.3.

    Add 20 μL Proteinase K Solution to each sample. Vortex to mix.

    1.4.

    Incubate at 55°C overnight in a shaking water bath.

    1.5.

    Centrifuge at maximum speed (≥4,000 x g) for 5 minutes to pellet undigested tissue

    debris and transfer 200 µL of the supernatant to the VWR 96-Well Deep Well Plate (1000 μL) without

    disturbing the undigested pellet

    List of materials to start to use OT-2

    2.

    It is necessary to have this equipment to use this pipeline.

    Opentrons Equipment List:

    AB
    OT-2
    OT-2 8-Channel Pipette P300
    OT-2 Magnetic Module GEN2Slot 7
    3.

    Tips & Labware:

    AB
    Four Opentrons 200µL Filter TipsSlot: 5,6,8,9
    Two Nest 1-Well Reservoirs (195 mL)Slot: 10 and 11
    VWR 96-Well Deep Well Plate (1000 μL)Slot 7 on the Magnetic module
    Two Nest Well plate ( 100 μL)Slot 2 and 3
    Two Nest 12 Well Reservoirs (15 mL)Slot 1 and 4

    Star OT-2 run

    4.

    The pipeline start on the step number 7 of the Mag-Bind® Blood & Tissue DNA HDQ 96 Kit Tissue Protocol

    Load the pipeline on the Opentrons app: Extracción.py

    The VWR 96-Well Deep Well Plate (1000 μL) labware can be download from: dx.doi.org/10.17504/protocols.io.dm6gp39njvzp/v1

    4.1.

    Order of the reagents on the Two Nest 12 Well Reservoirs (15 ml)

    Slot 1:

    ABCDEFGHIJKL
    Well 1Well 2Well 3Well 4Well 5Well 6Well 7Well 8Well 9Well 10Well 11Well 12
    11.5 mL of AL buffer11.5 mL of AL bufferMag-Bind mixed with HBQ buffer (20:340) in a total of 12 mLMag-Bind mixed with HBQ buffer (20:340) in a total of 12 mLMag-Bind mixed with HBQ buffer (20:340) in a total of 12 mL15 ml of VHB buffer15 ml of VHB buffer15 ml of VHB buffer15 ml of VHB buffer

    Slot 4:

    ABCDEFGHIJKL
    Well 1Well 2Well 3Well 4Well 5Well 6Well 7Well 8Well 9Well 10Well 11Well 12
    15 ml of VHB buffer15 ml of VHB buffer15 ml of VHB buffer15 ml of VHB buffer15 ml of SPM buffer15 ml of SPM buffer15 ml of SPM buffer15 ml of SPM bufferEllution Buffer